XW-12V5A8CH Battery Cell Charge and Discharge Grading Machine

Model: XW-12V5A8CH
Application: cylindrical lithium battery cell formation, capacity grading and consistency testing
Test range: 12V, 5A current, 8CH channel configuration
Cell type: cylindrical lithium cells and small-format lithium cells
Usage: cell production inspection, battery pack cell matching and quality traceability
Warranty: 1 year

General Details

XW-12V5A8CH Battery Cell Charge and Discharge Grading Machine is an 8CH battery cell charge and discharge grading machine for lithium cell capacity testing, cycling, aging and consistency evaluation. It is suitable for cylindrical cells and other small-format lithium cells when matched with the correct test fixture.

This model provides 12V test voltage, 5A current and 8CH independent channels. It is useful for battery laboratories, sample testing, pilot production and small-batch battery cell matching before pack assembly.

Machine Function

  • Run charge and discharge tests for lithium battery cells
  • Measure cell capacity, voltage behavior and cycle performance
  • Grade cells by consistency before battery pack assembly
  • Support aging, cycle testing and quality comparison
  • Record test data for analysis and production traceability

Key Features

  • 8CH independent test channels
  • 12V voltage and 5A current configuration
  • Computer software for test step setting and data management
  • Suitable for capacity sorting, charge-discharge testing and aging tests
  • Data export support for production records and quality reports
  • Customized fixtures available for cylindrical cell or other cell formats

Applications

  • Cylindrical lithium battery cell capacity grading
  • Battery cell charge-discharge cycle testing
  • Battery laboratory sample evaluation
  • Pilot line cell matching before pack assembly
  • Cell aging, quality comparison and R&D testing

Product Images

XW-12V5A8CH Battery Cell Charge and Discharge Grading Machine

XW-12V5A8CH Battery Cell Charge and Discharge Grading Machine

XW-12V5A8CH Battery Cell Charge and Discharge Grading Machine

XW-12V5A8CH Battery Cell Charge and Discharge Grading Machine

XW-12V5A8CH Battery Cell Charge and Discharge Grading Machine

XW-12V5A8CH Battery Cell Charge and Discharge Grading Machine

How to Select the Voltage and Current Range

The voltage and current range should be selected according to the cell type, nominal voltage, capacity and charge-discharge test rate. Higher voltage models are suitable when the test process or cell design requires a wider voltage window, while current selection should match the cell capacity and test standard.

Information Needed for Quotation

  • Cell type, size and capacity
  • Required voltage and current range
  • Channel quantity and daily testing capacity
  • Charge, discharge, rest and cycle test steps
  • Data report and export requirements
  • Fixture type and cell holder requirement

FAQ

What is this 8-channel grading machine used for?

It is used to charge and discharge battery cells, measure capacity and classify cells with similar performance before battery pack assembly or research testing.

Can it test cylindrical cells?

Yes. It can be used for cylindrical lithium battery cells when equipped with suitable fixtures. Other small-format cells may also be tested according to fixture design.

Can the software export test data?

Yes. Test data can be exported for production records, quality analysis, cycle comparison and cell matching.

Related Equipment

Request a Battery Cell Grading Solution

Send us your cell type, voltage range, current range, channel quantity and required testing process. XWELL can recommend a suitable battery cell charge and discharge grading machine.

Parameters
Device Model XW-12V5A8CH
Number of device channels 8​
Channel control mode More independent control of each channel, four-wire current and voltage acquisition
Channel characteristics The constant current source and constant voltage source adopt a double closed loop structure
Input Impedance ≥ 1MΩ​
Channel Interface The same port for charging and discharging, different ports can be customized
Channel Isolation Characteristics Channels share a common ground, fixtures of different channels cannot touch each other, and isolation customization is supported
Voltage range Charge 10mV-12V (device port)
Discharge 0V-12V (device port), supports forced discharge
Voltage accuracy ± 0.05% of FS, ambient temperature 15℃-35℃, other accuracy can be customized
Current range Charge 6mA-5A, dual range can be customized
Discharge 6mA-5A, dual range can be customized
Current accuracy ± 0.05% of FS, ambient temperature 15℃-35℃, other precisions can be customized
Power Single channel output power 60W
Stability 0.2% of FS
Time Current response time ≤ 10ms (90% range response time)
Data Recording Recording frequency 1Hz, customizable 10HZ
Data recording conditions Minimum time interval: 1S
Minimum voltage interval: 1mV or (related to the voltage range, 0.01% of the voltage range)
Minimum current interval: 1mA or (related to the current range, 0.01% of the current range)
Charge Charging Mode Constant current charging, constant voltage charging, constant current and constant voltage charging, constant power charging
Cut-off conditions Voltage, current, relative time, capacity, – △ V
Discharge Discharge mode Constant current discharge, constant power discharge, constant resistance discharge
Cut-off conditions Voltage, current, relative time, capacity, – △ V
Pulse mode optional Charge Constant current mode, constant power mode
Discharge Constant current mode, constant power mode
Minimum pulse width 5S or above is recommended
Cut-off conditions Voltage, relative time
DCIR DCIR (optional) Support custom points for DCIR calculation
Cycle Cycle Test 399 commands and 65535 cycles can be set
Nested loops Support 5 layers of loop nesting
Protect Software Protection Support power-off data protection (with UPS power supply, 100% guarantee of no data loss)
Support continued startup
Overvoltage protection, undervoltage protection, capacity protection, overcurrent protection, undercurrent protection
Reverse polarity protection
Data export method Excel 、 PDF
Cooling method Air Cooling
Equipment noise ≤ 50dB
Dimensions and weight Before packaging: 47x40x14cm (length, width and height); After packaging: 55x50x23cm (length, width and height); After packaging: 12kg
Communication method with host computer Network interface, through a switch, one computer can control multiple devices, and the network interface can be converted into a USB interface through a USB wired network card
System requirements Windows XP and above
Software Control Quantity Devices are connected via 485 or switches, and one computer can control more than 100 devices.
Temperature collection used with our temperature tester to collect battery temperature and generate a temperature-time curve. It can be used with our thermostat to control the thermostat temperature.
Other BMS communication protocol can be customized, other communication methods can be customized
Model List
Model Channel Voltage range Current range Single channel power
XW-5V3A8CH 8 Charge 0-5V, discharge 1.2-5V 0-3A 15W
XW-5V6A8CH 8 Charge 0-5V, discharge 1.2-5V 0-6A 30W
XW-5V10A8CH 8 Charge 0-5V, discharge 1.2-5V 0-10A 50W
XW-5V12A8CH 8 Charge 0-5V, discharge 1.2-5V 0-12A 60W
XW-6V5A8CH 8 Charge 0-6V, discharge 1.5-6V 0-5A 30W
XW-10V3A8CH 8 Charge 0-10V, discharge 1.5-10V 0-3A 30W
XW-10V5A8CH 8 Charge 0-10V, discharge 1.5-10V 0-5A 50W
XW-10V6A8CH 8 Charge 0-10V, discharge 1.5-10V 0-6A 60W
XW-12V5A8CH 8 Charge 0-12V, discharge 2-12V 0-5A 60W
XW-15V3A8CH 8 Charge 0-15V, discharge 2-15V 0-3A 45W
XW-15V5A8CH 8 Charge 0-15V, discharge 2-15V 0-5A 75W
XW-20V3A8CH 8 Charge 0-20V, discharge 2-20V 0-3A 60W
XW-25V3A8CH 8 Charge 0-25V, discharge 2-25V 0-3A 75W
XW-5V0.1A3A8CH 8 Charge 0-5V, discharge 1.2-5V 0-0.1A, 0.1-3A 15W
XW-5V0.5A6A8CH 8 Charge 0-5V, discharge 1.2-5V 0-0.5A, 0.5-6A 30W
XW-5V3A8CH-D 8 Charge 0-5V, discharge -5-5V 0-3A 15W
Video